Pulmonary Ventilator MV300

Price: 15000 MDL

Manufacturer: Mv-200

Product code: 78933

Description

Mechanical ventilators provide ventilatory support to critically ill patients with severe respiratory problems: ARDS, COPD, asthmatic status, morbid obesity, pulmonary hypertension, single lung, massive bacterial pneumonia, thrombembolism, lipid embolism, etc.).


Specification:


Mobile type, on a wheeled stand


Patient type - adul...

Mechanical ventilators provide ventilatory support to critically ill patients with severe respiratory problems: ARDS, COPD, asthmatic status, morbid obesity, pulmonary hypertension, single lung, massive bacterial pneumonia, thrombembolism, lipid embolism, etc.).


Specification:


Mobile type, on a wheeled stand


Patient type - adult Control/setting range Titral volume (inspiratory) ≥ 50-2000ml Peak flow 6-180 L/min Inspiratory pressure (Pinsp) 0-80 cmH2O Respiratory rate (RR) 0-80rpm Adult Inspiratory time (Ti) 0. 2 - 5s Expiration time I:E ratio 1:9 to 4:1 Pause on inhale Pause on exhale FiO2, % 21-100 Manual breathing button PEEP/CPAP 0-50 cmH2O Pressure support (Psupp) 0-100 cmH2O Nebulizer Pressure -0.5-15 cmH20 Flow 0.5-20 L/min Button 100 % O2 Control panel lock


Operating modes


 A/C mode VA/C Breathing volume P-A/C breathing pressure SIMV mode V-SIMV breathing volume P-SIMV breathing pressure CPAP mode CPAP/PEEP mask CPAP/PEEP mode Apnea-backup mode Intelligent or combined breathing control modes Non-invasive ventilation (NIV) via mask APRV bilevel mode Parameters monitored/displayed Cstat (static compliance), Static Compliance Cdyn (dynamic compliance), Dynamic compliance Transpulmonary Pressure Airway Resistance PEEP Pressure Total Volume ,FiO2, ETCO2 ,Respiratory Rate ,Inspiratory Time,Expiratory Time, I Rate: E ,Spontaneous minute volume ,Patient alarms FiO2 high/small ,Minute volume high/small ,Inspiratory pressure high/small ,PEEP high ,Apnea ,Continuous high pressure/occlusion ,Leaking gas mixture in circuit ,Disconnected breathing circuit ,Equipment alarms, Gas supply failure ,Electrical supply failure ,Battery low ,System error, Self-diagnostics, TFT LCD display, Touchscreen ,Size = 15 inch Brightness adjustment,


Other requirements


Removable one-way valves, O2 sensor, Air source "Integrated in the device, turbine type, electrically operated", O2 source Possibility to connect to central O2 source Permissible working pressure 2.8-6.0 bar Power supply Mains 220 V, 50 Hz , Internal rechargeable battery, Battery operating time ≥ 2 h


Accessories


Reusable adult respirator type 2 set. Reusable adult respirator 1 set. ETCO2 monitoring via reusable tracer. Include all the necessary components to enable ETCO2 monitoring in a minimum of 40 patients, 40 patients SpO2 monitoring. Humidifier VADI VH-2000 with servo control Humidifier with temperature maintenance and automatic air humidification, Ventilator compatible, Minimum two operating modes: invasive and non-invasive,Reusable adult humidification chamber 2 pcs.


Adult disposable antibacterial filters 100 pcs. Adult reusable flow sensor 2 pcs. Adult reusable test screen 1 pc. M01=4 castors,= 2 castors with brake,Articulated arm for fixing breathing hoses, Bracket for fixing/attaching electrical cables, air hose, oxygen for transport, storage,


Handle for transport


Product and supplier requirements


User manual (in state language), Service manual, technical maintenance of the device with access passwords, "Training for bioengineers (device maintenance, maintenance, calibration, etc.), Training for user (working principle, ventilation regimes, maintenance, other)"